Onset of Atrial Fibrillation: Mode Switch
Report:Sinus rhythm 76/min 1 Atrial-sensing ventricular pacemaker 3 SVEB, onset of atrial fibrillation 4 Atrial fibrillation with ventricular response 118-124/min 1 Pacemaker inhibition 0 Incomplete left bundle branch block 1 Comment:At first, the p
Slowing Rate with Battery Depletion
Report:Atrial fibrillation 1 Pacemaker rhythm 60/min 2 Possible battery depletion 7 Comment:Battery depletion is usually programmed to manifest itself by slowing to a pre-set rate, in this case 60/min. This patient’s potential problem was in fact dia
Sensing and not Sensing Atrial Fibrillation
Report:AV sequential pacemaker rhythm 60-75/min 5 Atrial fibrillation 5 Comment: For much of the trace the pacemaker is firing through both barrels, unaware that AF precludes any atrial capture. This wastes the battery. On the other hand, having a rapi