Battery Depletion
Report: Atrial fibrillation 1 Third degree AV block 4 Pacemaker rhythm 44/min 1 Probable battery depletion 4 Comment: Most units are programmed to slow the basic rate as the battery becomes depleted; numbers vary. It is not certain, however, that batt
Left Ventricular Epicardial Pacemaker
Report:AV pacemaker rhythm 81/min 2 Unusual axis ? left ventricular electrode 8 Comment:The pacemaker has 100% atrial and ventricular capture. The unusual RAD in the frontal plane and RV1 suggest a left ventricular electrode. This may be deliberate or
End-of-Life Pacemaker Function
Report:Sinus rhythm 75 – 77/min 1 Third degree AV block 2 Pacemaker rhythm 31/min 2 Probable battery depletion 5 Comment:Pacemaker interrogation revealed “end-of-life” battery function, set to bipolar VVI mode at 63/min, but obviously firing at half t
Onset of Atrial Fibrillation: Mode Switch
Report:Sinus rhythm 76/min 1 Atrial-sensing ventricular pacemaker 3 SVEB, onset of atrial fibrillation 4 Atrial fibrillation with ventricular response 118-124/min 1 Pacemaker inhibition 0 Incomplete left bundle branch block 1 Comment:At first, the p
Atrial Pacemaker: Intermittent Failure to Sense
Report: Adler Roy 18/4/99 11065929Sinus rhythm 66/min 1 Atrial pacemaker discharge 53/min Intermittent failure to sense 2 Pacemaker capture beats 2 Phasic aberrant RBBB conduction 2 Left atrial abnormality (LAA) 1 Anterior infarction, indeterminate
Single Failure to Capture
Report:Sinus rhythm 63/min 2 Complete AV dissociation 4 Pacemaker rhythm 80/min 2 Single failure to pace 4 Comment:AV block cannot be accurately diagnosed here: the atrial (sinus) rate is too slow and the ventricular (pacemaker) rate too fast. Just be
Symptomatic Bigeminy Due to Trigeminy
Report: Sinus rhythm 80/min 1 Blocked SVEBs, trigeminal 8 A-sensing ventricular pacing throughout 1 Comment: Not surprisingly, the pacemaker was blamed. It was programmed to pace at 60/min, with a sleep mode of 50/min during night hours. There was no
Retrograde Wenckebach and Reentry
Report:Pacemaker rhythm 92/min 1 Retrograde Wenckebach conduction 4 Wenckebach sequences terminated by reentry (echo) beats of pacemaker origin 4 Non-specific T wave changes 1 Comment:The pacemaker is in its rate-responding mode at 92/min, with Wenck
Allorhythmic Failure to Pace & Sense
Report: Sinus rhythm 59/min 1 First degree AV block 1 Non-specific ST/T changes 1 Demand pacemaker 75/min 2 Trigeminy 1 Intermittent failure to pace 1 Possible intermittent failure to sense 1 Comment: Of every three pacemaker spikes, only one - the
Atrial Pacemaker: Complete Failure to Sense
Report:Sinus/atrial pacemaker rhythm 58-72/min 5 Atrial pacemaker: failure to sense 3 Borderline low voltage in frontal leads 1 Nonspecific ST/T changes, V2-5, consistent with ischæmia 1 Comment:The atrial pacing rate is much slower than the sinus rhy